Dirty
Films
- Verify
Accuracy of Processor Rollers are they Clean & Not Dirty?
- Verify that
Processor is Cleaned Monthly as Recommended.
- Verify and
Repair Any Questionable Water Problems within the Processor.
- Verify
Integrity of Calibration of Processor is Developer Temperature Accurate?
- Verify and
Replace All Depleted Developer Chemistry In and Under the Processor.
- Verify and
Replace All Questionable or Contaminated Chemistry. Developer/fixer.
- Verify
Accuracy of Chemistry has it been Properly Mixed? When was problem Noted?
- Verify
Accuracy and Repair any Circulation Problems within the Processor. Check Pumps.
- Verify that
Developer In-line Filters are Replaced as recommended by the manufacturer.
- Verify that
All Processor Chemicals are Fresh and Rotated or Replaced Monthly.
- Verify that
all Processor Drain Lines are Replaced as Needed or Annually.
Stains
on Films
- Verify
Accuracy of Chemical Replenishment Rates.
- Yellow =
Old Exhausted Fixer
- Diachronic
= Bad Chemistry
- Green =
Insufficient Fixing
- Metallic
= Oxidized Developer Possibly a
water/Fixer/Drain Back-Up Contaminating Developer.
- White-Crystalline
= Defective Milky Fixer, Developer
Splashed into Fixer, Insufficient Rinsing.
- Grit =
Dirty Water or Dirt in Dryer Assembly
(See
Chemical Problems Section)