Increased Base Fog
Darkroom:
- Verify
Accuracy of Developer Replenishment Rates.
- Verify
Accuracy of Developer Temperature is it too High?
- Verify
Amount of Starter Added to Developer in the processor.
- Verify
Accuracy of Fixer levels in the processor is it set low?
- Verify that
Developer has not been Contaminated by Fixer.
- Verify that
Processor's Rollers are not Dirty or Worn.
- Verify
Integrity of Safelight Check Filter and Bulb. Is it proper type for Film?
- Verify
Integrity of Darkroom, Check Exhaust Vents, Verify no Light Leaks or Heat.
- Verify
Accuracy of Shipping & Handling of Film to Facility. Check Expiration Date.
- Verify
Accuracy and Quality Film Vendor is it a name brand?
- Verify Film
is kept away from Chemicals, Light, Scatter Radiation & Heat.
- Verify
Location of Film Bin from X-Ray Room. Adequate Shielding?
- Verify
Location and Storage of Film in Cassettes. Where are Cassettes Stored in Room?
- Verify
Vendors Facility. Film Stored in a Air Conditioned Area or Hot Warehouse?
- Verify
Storage of Film Received from Vendor. How was it Delivered in Hot Trunk of Car?
X-Ray:
- Verify
Location of Cassettes. Where are they Placed during Exposures?
- Verify
Proper Shielding in X-Ray Room. Check for Scatter Radiation.
- Verify that
all loaded cassettes are not stored near Heat, Sunlight, or Radiation.