High
Density or Increased Contrast
Darkroom:
- Verify
Accuracy of Developer Temperature is it unstable or set too high?
- Verify
Accuracy of Developer was it Properly mixed or Off with too little water.
- Verify
that Developer Replenishment Rates are Properly Set and not set too High.
- Verify
If Proper Amount of Starter was added to the Developer.
- Verify
Incoming Water Temperature that it is cold and not too Hot.
- Verify
that Incoming Water Valve is On and Water is Circulating Properly.
- Verify
Films Quality and Proper Storage, Check Expiration Date.
- Verify
that Film has not been exposed to Chemical Fog or Excessive Heat.
- Verify that
Film has not been exposed to Scatter Radiation or Light.
- Verify
Integrity of Film, Cassettes & Screens.
X-Ray:
- Verify
Proper Exposure Factors. (kVp/ Ma/Timer set too High/Over Exposure?)
- Verify
Proper Focal Distance. (Too Short Focal Distance?)
- Verify
Calibration of System. (kVp/Ma/Timer Accurate?)
- Verify if
Power deviates During the Day or while Making Exposures.
- Verify
Accuracy and Proper Positioning of Patient.
- Verify
Accuracy of AEC System and Incorrect Centering of Patient to Chamber.