Low Density or Reduced Film Speed
Darkroom:
- Verify
Accuracy of Developer Temperature is it Properly Set or Undeveloped/ too Low?
- Verify
Accuracy of Thermometer when was the last time it was checked?
- Verify
Proper Amount of Chemistry in Tanks is Chemistry Low? Exhausted Developer?
- Verify
Accuracy/Calibration of Chemical Replenishment Rates are they set too Low?
- Verify
Accuracy of Developer is it Fresh and Properly Mixed.
- Verify
Accuracy of Chemical Recirculation in Processor, Check Pumps.
- Verify
Accuracy of Developer that it is not Contaminated by Fixer or Water.
- Verify that
Processor Main Drain is Properly Working, no Plumbing Issues or Back-Ups.
- Verify
Accuracy of Developer is it Fresh or weak, Light or Dark, has it oxidized?
- Verify
Accuracy and Calibration of Film to Screens are you using the Proper Film?
- Verify
Quality of Film is it a Name Brand or Generic, Check Freshness and Expiration Date.
- Verify
Accuracy and Quality of Cassette Screens are they Weak or Questionable?
- Verify that
all Cassette Screens are the Same. Check dates and they Cleaned Frequently?
- Verify
Accuracy of Film Drive Path in Processor is it Questionable? Belt ever Changed?
X-Ray:
- Verify
System Calibration. (kV/Ma/Timer/Focal Distance Off?)
- Verify
Proper Line Voltage to System no Transients or Power Surges.
- Verify
Alignment of Light to Radiation on System. Call for Service.
- Verify
Accuracy of AEC/Automatic Exposure Control and Calibration.
- Verify
Proper Centering of Patient to Bucky/AEC Chambers.
- Verify
Distance to Grid/Bucky Possibly out of Grid Radius?
- Verify
Accuracy of System. How Old Is Tube/Transformer/ Questionable?